title: "Diabetes app alarms can fall silent on Android 15, Spain warns"
description: "Spain’s medicines agency says MiniMed Go version 1.3.0 fixes the fault. The sound can fail when the app runs in the background after a phone restart."
url: https://thehealthinsights.eu/news/minimed-go-app-android-alarm-fault
markdown_url: https://thehealthinsights.eu/news/minimed-go-app-android-alarm-fault.md
published: 2026-11-29T02:59:31.181Z
section: "Digital Health"
countries: ["Spain", "Germany", "United Kingdom", "Switzerland"]
topics: ["Diabetes", "Medical devices"]
authors: ["The Health Insights Desk"]
editor: "The Health Insights Desk"
publisher: "The Health Insights"
language: en-GB
image: https://thehealthinsights.eu/api/og/news/minimed-go-app-android-alarm-fault/16x9
image_credit: "Generated artwork"
copyright: "© 2026 The Health Insights"
license: https://creativecommons.org/licenses/by/4.0/
reuse_policy: https://thehealthinsights.eu/reuse
cite_as: "The Health Insights Desk. \"Diabetes app alarms can fall silent on Android 15, Spain warns\". The Health Insights, 29 Nov 2026. https://thehealthinsights.eu/news/minimed-go-app-android-alarm-fault"
---

# Diabetes app alarms can fall silent on Android 15, Spain warns

*Spain’s medicines agency says MiniMed Go version 1.3.0 fixes the fault. The sound can fail when the app runs in the background after a phone restart.*

By [The Health Insights Desk](https://thehealthinsights.eu/contributors/the-health-insights-desk), edited by [The Health Insights Desk](https://thehealthinsights.eu/contributors/the-health-insights-desk). Published 29 Nov 2026, 03:59 CET. [Digital Health](https://thehealthinsights.eu/section/digital-health). Countries: Spain, Germany, United Kingdom, Switzerland. Topics: [Diabetes](https://thehealthinsights.eu/topic/diabetes), [Medical devices](https://thehealthinsights.eu/topic/medical-devices).

**Why it matters:** Spain’s medicines agency asks healthcare professionals to tell MiniMed Go users about the fault and recommend the app update to those on Android 15.

Spain’s medicines agency, AEMPS, warned that the MiniMed Go diabetes app could stop sounding glucose alerts after a phone restart in a [26 November notice](https://www.aemps.gob.es/informa/la-aemps-informa-de-un-problema-en-la-aplicacion-para-la-diabetes-minimed-go-en-dispositivos-android-15-que-afecta-a-las-alertas-sonoras/).

The sound can fail when the app runs in the background on phones with Android 15, the agency said. People who rely on sound while asleep or away from their screen may miss a glucose alert. AEMPS said delayed intervention could allow low or high blood glucose to worsen.

**How it works:** MiniMed Go combines readings from a continuous glucose sensor with records of insulin given through the InPen smart pen, AEMPS said. In Spain, the agency names Simplera and Instinct Go as compatible sensors. MiniMed identifies the Android app as model MMT-8601 in its [field safety notice](https://www.aemps.gob.es/informa/notasInformativas/productosSanitarios/2026/NI-PS-49/147151-FSN-ES.pdf).

**Yes, but:** Visual alerts and vibration still work, MiniMed said in its notice, which recorded no complaints about the fault. AEMPS gave no count of affected users. The notice covers MiniMed Go versions below 1.3.0 on Android 15, and AEMPS said users on Apple devices need take no action.

**What AEMPS asks:** The agency asked healthcare professionals to tell MiniMed Go users about the fault and recommend the update to those on Android 15. It directed those users to Google Play for version 1.3.0 or later. It also pointed users to MiniMed’s technical support line and asked anyone aware of a device incident to report it through [NotificaPS](https://notificaps.aemps.es/).

Germany’s medical devices regulator [posted the manufacturer’s notice](https://www.bfarm.de/SharedDocs/Kundeninfos/DE/09/2026/52787-26_kundeninfo_de.html) on 18 November. The UK’s medicines and devices regulator [listed it](https://www.gov.uk/drug-device-alerts/field-safety-notices-16-to-20-november-2026) on 24 November. The Swiss patient association diabètesuisse [shared the warning](https://www.diabetesuisse.ch/personnes-touchees-et-proches/information/actualite/news-detail-fr/news/sicherheitshinweis-minimed-gotm-app-auf-android-os-15) on 17 November and pointed users to the same update.

AEMPS issued [a separate notice in July](https://www.aemps.gob.es/informa/la-aemps-informa-de-un-posible-fallo-en-la-aplicacion-minimed-go-para-la-diabetes-que-afecta-al-funcionamiento-de-las-alertas-nocturnas/) about two overnight alerts. The app could fail to send a missed-dose reminder or a prompt to correct high glucose when its day and night mode was off. AEMPS identified version 1.2.0 as the fix for that fault.

### What's next

On 26 November, AEMPS said Medtronic Ibérica was emailing affected users in Spain and would follow up with an in-app message about the update. The company will also alert Android 14 users in case they later upgrade their phones to Android 15.

## Sources

- Agencia Española de Medicamentos y Productos Sanitarios, [La AEMPS informa de un problema en la aplicación para la diabetes MiniMed Go en dispositivos Android 15 que afecta a las alertas sonoras](https://www.aemps.gob.es/informa/la-aemps-informa-de-un-problema-en-la-aplicacion-para-la-diabetes-minimed-go-en-dispositivos-android-15-que-afecta-a-las-alertas-sonoras/), 2026-11-26 (primary)
- MiniMed Iberia, [Error de sonido de las alertas en segundo plano de la aplicación MiniMed Go para el SO 15 de Android](https://www.aemps.gob.es/informa/notasInformativas/productosSanitarios/2026/NI-PS-49/147151-FSN-ES.pdf), 2026-11-26 (primary)
- Bundesinstitut für Arzneimittel und Medizinprodukte, [Dringende Sicherheitsinformation zu Minimed Go Application von Medtronic Minimed](https://www.bfarm.de/SharedDocs/Kundeninfos/DE/09/2026/52787-26_kundeninfo_de.html), 2026-11-18 (primary)
- Medicines and Healthcare products Regulatory Agency, [Field Safety Notices: 16 to 20 November 2026](https://www.gov.uk/drug-device-alerts/field-safety-notices-16-to-20-november-2026), 2026-11-24 (primary)
- diabètesuisse, [Avis de sécurité : application MiniMed Go™ sur Android OS 15](https://www.diabetesuisse.ch/personnes-touchees-et-proches/information/actualite/news-detail-fr/news/sicherheitshinweis-minimed-gotm-app-auf-android-os-15), 2026-11-17 (stakeholder)
- Agencia Española de Medicamentos y Productos Sanitarios, [La AEMPS informa de un posible fallo en la aplicación MiniMed Go para la diabetes que afecta al funcionamiento de las alertas nocturnas](https://www.aemps.gob.es/informa/la-aemps-informa-de-un-posible-fallo-en-la-aplicacion-minimed-go-para-la-diabetes-que-afecta-al-funcionamiento-de-las-alertas-nocturnas/), 2026-07-16 (primary)

## Citation

Cite as: The Health Insights Desk. "Diabetes app alarms can fall silent on Android 15, Spain warns". The Health Insights, 29 Nov 2026. https://thehealthinsights.eu/news/minimed-go-app-android-alarm-fault

APA: The Health Insights Desk. (2026, November 29). Diabetes app alarms can fall silent on Android 15, Spain warns. The Health Insights. https://thehealthinsights.eu/news/minimed-go-app-android-alarm-fault

Vancouver: The Health Insights Desk. Diabetes app alarms can fall silent on Android 15, Spain warns \[Internet\]. The Health Insights; 2026 Nov 29. Available from: https://thehealthinsights.eu/news/minimed-go-app-android-alarm-fault

© 2026 The Health Insights. Text licensed under [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/): reuse it with credit to The Health Insights and a link to this story. Images credited to others keep their own licence. [Reuse and citation](https://thehealthinsights.eu/reuse). Every story is approved by a named editor before publication; corrections are noted at the foot of the story ([editorial standards](https://thehealthinsights.eu/standards)).

Canonical URL: https://thehealthinsights.eu/news/minimed-go-app-android-alarm-fault